Setting CPAP humidification: a step by step guide
Three settings not to confuse
A dry nose, a sore throat and droplets in the mask all come from the same place: humidification. But three different settings control it, and mixing them up means going round in circles.
| Setting | What it controls | Values |
|---|---|---|
| Humidity Level | How much water is added to the air | Off, or 1 to 8 |
| Tube Temperature | How warm the air is when it reaches the mask | Off, automatic, or 16 to 30 degrees |
| Climate Control | How the two above are managed | Automatic or manual |
Tube temperature and Climate Control are only available with ClimateLineAir heated tubing. Without it, only the humidity level exists. Our heated CPAP tube guide explains what that tubing brings.
Which level should you choose?
There is no universally correct setting. It depends on your bedroom, the season and your nose. The method that works is to start from the symptom.
- Dry nose, sore throat, nosebleeds: raise the humidity level
- Droplets in the mask or tubing: lower the humidity level, or raise the tube temperature
- Very dry mouth on waking: this is not the humidifier. Air is escaping through your mouth overnight
Change one step at a time and give it two or three nights. Jumping from 3 to 7 swaps one problem for another without ever finding the right point.
Setting humidification on a ResMed AirSense 11
Everything happens in MY OPTIONS, on the touch screen.
Automatic mode, the one to prefer
Climate Control on Auto is the setting ResMed recommends, and the one shipped by default. The device holds the tubing at 27 degrees and 85 percent relative humidity, and protects against condensation. There is nothing for you to set.
- From the Home screen, tap MY OPTIONS.
- Tap Climate Control.
- Tap Auto, then OK.
Manual mode, if automatic does not suit you
In manual mode you set temperature and humidity separately. You gain control, but protection against condensation is no longer guaranteed.
- MY OPTIONS, then Climate Control, then Manual, then OK.
- Go back into MY OPTIONS and tap Humidity Level. Pick 1 to 8, then OK.
- Tap Tube Temp. Pick a value between 16 and 30 degrees, then OK.
Setting humidification on a ResMed AirSense 10
The AirSense 10 runs on the dial: turn to move, press to confirm.
- From the Home screen, turn the dial to My Options, then press.
- Turn to Humidity Level, then press.
- Turn to pick a level from 1 to 8, or Off.
- Press to save.
Condensation in the tubing
You hear water gurgling, or you get droplets on your face. That is condensation: humid air cools as it travels down the tubing and the water settles on the walls.
Three levers, in this order:
- Raise the tube temperature, so the air stays warm all the way to the mask
- Lower the humidity level by one step
- Route the tubing under the duvet to benefit from body heat, and move the bed away from a cold window
A correctly set heated tube solves the problem in the large majority of cases.
The heated tube, by device:
Three common mistakes
Filling the tub with hot water
ResMed forbids it: the water must be at room temperature. Hot water can send excessively warm air all the way to the mask.
Using the wrong water
ResMed recommends tap water that has been boiled and cooled, distilled water, or bottled water with a low mineral content. It rules out non-sterile household demineralised water, the kind sold for irons. In every case the water must be at room temperature and must never go past the maximum level mark, which corresponds to 380 ml.
Leaving the humidifier on while running on battery
The humidifier is by far the biggest power draw on the device. On a CPAP battery, turn it off to double your runtime.
Read more
- CPAP humidifier: the complete guide
- The heated CPAP tube
- Setting your CPAP ramp
- CPAP side effects and how to fix them
All our settings tutorials
Each tutorial gives the procedure for both the AirSense 11 and the AirSense 10.