Cleaning your CPAP machine: the full routine

5 min read

The routine, in two parts

Looking after a CPAP comes down to two habits. A daily one that takes a minute, and a weekly one that takes ten. The rest is periodic inspection.

How often What you do
Every day Empty the tub, wipe it, refill with the water its manual specifies
Every week Wash the tub and tubing, wipe the outside of the device
Every six months Replace the air filter, inspect the tubing and tub
First of all: unplug the device and let the humidifier cool for at least ten minutes. The heater plate stays hot after shutdown. Never immerse the device, the AC adaptor or the power cord in water.

What to wash with, and above all what not to

ResMed is explicit on this point, and this is where the real damage happens.

  • Allowed: warm water and a mild dishwashing liquid. For the tub, a solution of one part white vinegar to nine parts water at room temperature
  • Forbidden: bleach, chlorine, aromatic based solutions, moisturising or antibacterial soaps, scented oils
  • Also forbidden: washing above 55 degrees

These products damage the materials, reduce humidifier performance and void the warranty. An antibacterial soap is not a bonus precaution, it is a mistake.

Cleaning a ResMed AirSense 11

Taking it apart

  1. Hold the tub at the top and bottom, press it gently and pull it away from the device.
  2. Open the tub and discard any remaining water.
  3. Pinch the cuff of the air tubing and gently pull it away from the device.
  4. Hold the cuff of the tubing and the swivel of the mask, then gently pull apart.
  5. Locate the outlet connector inside the device and release it by pressing the clip firmly.
  6. Remove the outlet connector by pulling it out through its socket at the rear of the device.

Washing

Every day, empty the tub, wipe it thoroughly with a clean disposable cloth and let it dry out of direct sunlight.

Every week, wash the tubing, the tub and the outlet connector in warm water with a mild dishwashing liquid. Rinse thoroughly, then dry out of direct sunlight and away from any heat source. Wipe the exterior of the device with a dry cloth.

On the AirSense 11, the tub and outlet connector are dishwasher safe. The tubing never is.

Putting it back together

  1. Hold the outlet connector with the seal pointing left and the clip pointing forward.
  2. Align it correctly and insert it into the socket.
  3. Check that it is fully seated.
  4. Connect the tubing firmly to the air outlet at the rear of the device.
  5. Open the tub and fill it to the maximum water level mark with water at room temperature: drinking or distilled water (the HumidAir 11 tub sold in Europe is the cleanable one), never the demineralised water sold for steam irons.
  6. Close the tub and insert it into the side of the device.
  7. Connect the free end of the tubing firmly to the assembled mask.

Cleaning a ResMed AirSense 10

Taking it apart

  1. Hold the tub at the top and bottom, press it gently and pull it away from the device.
  2. Open the tub and discard any remaining water.
  3. Hold the cuff of the tubing and gently pull it away from the device.
  4. Hold the cuff of the tubing and the swivel of the mask, then gently pull apart.

Washing

Every week, wash the tub and tubing in warm water using only a mild detergent. Rinse thoroughly, let dry out of direct sunlight and heat, then wipe the exterior of the device with a dry cloth.

Important difference from the AirSense 11: on the AirSense 10, neither the tub nor the tubing goes in a dishwasher or washing machine. This is the most common mistake among people switching devices.

Putting it back together

  1. Connect the tubing firmly to the air outlet at the rear of the device.
  2. Open the tub and fill it with distilled room temperature water up to the maximum mark.
  3. Close the tub and insert it into the side of the device.
  4. Connect the free end of the tubing to the assembled mask.

The inspection everyone forgets

Cleaning is not enough, you also have to look. Once a month, take thirty seconds.

  • The tub: replace it if it leaks, if it is cracked, cloudy or pitted. Replace it too if the seal is cracked or torn
  • White deposits: those are minerals. Remove them with a solution of one part white vinegar to nine parts water, then rinse with clean water
  • The tubing: replace it at the first hole, tear or crack. A pinhole leak in the tubing shows up as leaks you will hunt for in vain around the mask

What this page does not cover

The mask has its own routine, different and more frequent. It is covered in our guide on cleaning your CPAP mask.

The air filter is not washable and not reusable, it gets replaced. That is the subject of our dedicated page on filter replacement.

The inside of the device cannot be cleaned at home:

CPAP disinfection service - Workshop decontamination of your device
View

Read more

All our settings tutorials

Each tutorial gives the procedure for both the AirSense 11 and the AirSense 10.

Frequently asked questions

Can I put my tub in the dishwasher?

It depends on your device. On the AirSense 11, the tub and outlet connector are dishwasher safe. On the AirSense 10 the manual forbids it outright, for both the tub and the tubing. In both cases the tubing never goes in a dishwasher.

Does white vinegar damage the tub?

No, provided you respect the dilution. ResMed recommends one part vinegar to nine parts water, at room temperature. That is the stated method for removing white mineral deposits. Always rinse with clean water afterwards.

Can I use antibacterial soap?

No, and it is counterintuitive. ResMed forbids antibacterial and moisturising soaps, bleach, chlorine, aromatic solutions and scented oils. They degrade the materials and can void the warranty. A mild dishwashing liquid is enough.

How often should the tubing be replaced?

There is no mandated interval, condition decides. Replace it as soon as it shows a hole, a tear or a crack.

Do I really have to empty the tub every morning?

Yes, it is the single most important habit. Water sitting all day in a warm tub is an ideal breeding ground. Empty it, wipe it with a clean cloth, let it dry, and refill just before the next night.

Are ozone cleaning devices recommended?

No, and the most important reason concerns your health before your device. The FDA states it has received reports of asthma attacks, headaches and shortness of breath after using ozone devices, and notes that the ozone concentration inside the equipment can remain above safety limits for several hours after the cycle ends, while that is the circuit you then breathe through all night. Both ozone and UV light can break down tubing and masks, and the FDA has received no data showing that either one cleans the inside of CPAP tubing. ResMed also excludes damage caused by an ozone device from its warranty for machines sold since 1 February 2020. The manual cleaning described here remains the manufacturer approved method.